SurePette® Non-autoclavable Adjustable Pipette
Product Description
CATALOG
CATALOG NO. | PRODUCT NAME | CERTIFICATE |
ABT-SPA-10 | SurePette Adjustable Pipette 0.5 - 10 μl | CE |
ABT-SPA-20 | SurePette Adjustable Pipette 2 - 20 μl | CE |
ABT-SPA-50 | SurePette Adjustable Pipette 5 - 50 μl | CE |
ABT-SPA-100 | SurePette Adjustable Pipette 10 - 100 μl | CE |
ABT-SPA-200 | SurePette Adjustable Pipette 20 - 200 μl | CE |
ABT-SPA-1000 | SurePette Adjustable Pipette 100-1000 μl | CE |
ABT-SPA-5000 | SurePette Adjustable Pipette 1000-5000 μl | CE |
INTRODUCTION
1. SUREPETTE® Series Pipettes include Adjustable Volume Pipettes and Fixed Volume Pipettes, covering a volume range from 0.5 μL to 5 mL;
2. SUREPETTE® Series Pipettes aspirate and dispense volumes of liquid precisely and safely;
3. A volume indicator makes it easy to set volume and to eliminate calculation errors;
4. A tip ejector button simplifies the discarding of the tip and makes the operation safe;
5. Each SUREPETTE® is factory calibrated and carefully checked gravimetrically using distilled water and an analytical balance.
PIPETTING
1. Attach a new disposable tip to the tip holder. Press only enough to make a positive airtight seal. This avoids leaks or poor accuracy;
2. Pre-rinse the tip by aspirating the first volume of liquid and then dispensing it back or to waste;
3. Press the push button to the first stop. This part of the stroke is the volume displayed on the volume indicator;
4. Holding SUREPETTE® vertically, immerse the tip into the liquid to a depth of 2 – 4 mm;
5. Wait 2 – 3 seconds. Allow the push button to return slowly and smoothly to the top position. DO NOT let it snap up;
6. Withdraw the tip from the liquid after the full volume of sample is drawn into the tip, touching against the edge of the container to remove excess;
7. To dispense sample, place the tip end against the inside wall of the recipient vessel at an angle of 10°to 40°. Depress the push button slowly and smoothly to the first stop. Wait 1 – 3 seconds. Then press the push button to the second stop (bottom of stroke), expelling the residual liquid from the tip;
8. Withdraw SUREPETTE® from the vessel carefully. Allow the push button to return to the top position;
9. Discard the tip by depressing the tip ejector.
SPECIFICATIONS
Model & Volume Range | Normal Volume | Accuracy | Precision |
(mean error) | (repeatability) | ||
μl | % | % | |
SPA - 10 0.5~10 μl | 1 * | ±12 | ≤4.0 |
5 | ±6.4 | ≤3.0 | |
10 | ±3.2 | ≤2.0 | |
SPA -20 2~20 μl | 2 | ±10 | ≤3.0 |
10 | ±3.2 | ≤2.0 | |
20 | ±3.0 | ≤1.5 | |
SPA - 50 5~50 μl | 5 | ±6.4 | ≤3.0 |
25 | ±3.0 | ≤2.0 | |
50 | ±2.4 | ≤1.5 | |
SPA - 100 10~100 μl | 10 | ±3.2 | ≤3.0 |
50 | ±2.4 | ≤1.5 | |
100 | ±1.5 | ≤1.0 | |
SPA - 200 20~200 μl | 20 | ±3.0 | ≤1.5 |
100 | ±1.5 | ≤1.0 | |
200 | ±1.5 | ≤1.0 | |
SPA - 1000 100~1000 μl | 100 | ±1.5 | ≤1.0 |
500 | ±1.5 | ≤0.5 | |
1000 | ±1.2 | ≤0.5 | |
SPA - 5000 1000~5000 μl | 1000 | ±1.5 | ≤0.5 |
2500 | ±1.0 | ≤0.5 | |
5000 | ±1.0 | ≤0.5 |
Note: The above information is for reference use only. Please refer to the product insert provided with the products before use.